Defibrillator-Sanjeevani 1001-1005

Defibrillator

Sanjeevani 1001-1005
  

Problem - Solutions List

  • Display colour changes
    1. Solution :FPC cable connecting LCD to PCB 13002 (ident J14) is not connected properly. Replace the cable or connect it properly.

      Details :Refer reference document for making changes in the unit and Image 2

  • ECG traces are flat
    1. Solution :Check crimping and continuity of ECG connector from PCB 9014 (ident CN2) to the side panel of the unit.

      Details :Refer Cable SP1 and Image 1

    2. Solution :Change PCB 9014 - ECG amplifier.

      Details :Refer reference document for making changes in the unit and Image 1

  • Encoder broken
    1. Solution :Check the encoder, If broken replace it .(Encoder - encoder pec16-4220f-s0012; 20MM SHAFT w/switch).

      Details :Refer Cable FP3

  • It does not give 'PAPER OUT' message
    1. Solution :Flex cable of the printer not connected properly to PCB 12027 (ident J2). Check and connect it properly.

      Details :Refer Document A1

    2. Solution :If Flex cable is connected properly then check the printer (check with new printer).

      Details :Refer Document A1

    3. Solution :Change PCB 12027 - Printer PCB.

      Details : Refer reference document for making changes in the unit and Refer Document A1

  • Keyboard not working
    1. Solution :Replace keyboard and check it. The FRC cable is connected to PCB 13002.

      Details :Refer Document A2

    2. Solution :Replace PCB 13002 and check it. The FRC cable is connected to PCB 13002.

      Details :Refer reference document for making changes in the unit and Image 2

  • Keys on keyboard not working
    1. Solution :Replace keyboard. The FRC cable of the keyboard is connected to PCB 13002.

      Details :Refer Document A2

  • Mains-Batt indication is not glowing
    1. Solution :Check continuity of 3 pin Relimate connector on PCB 13002 (ident J5) to 3 pin Relimate connector on PCB 9013 (ident CN12), if faulty replace it.

      Details :Refer Cable CC4

    2. Solution :If the cable CC4 is not faulty then replace keyboard.

      Details :Refer Document A2

  • Manual check process is not completed
    1. Solution :Check continuity of U plate connection from Apex handle to PCB 11006 (ident CON6), If faulty replace it.

      Details :Refer Cable FP8 and Image 6

    2. Solution :Check continuity of U plate connection from Sternum handle to PCB 12026 (ident CON5)), If faulty replace it.

      Details :Refer Cable FP7 and Image 6

  • Manual check process is not completed (Only "Place pads in cradle" message is displayed)
    1. Solution :Check if handle is placed properly in the cradle along with the adult plate. Remove the handle and place it in the cradle again.

      Details :NA

    2. Solution :Check POT (ident Pot P1) on PCB 9013 to 3.28v with respect to ground of (ident CN3) connector. Handle should be placed in the cradle.

      Details :Image 1

  • QRS Volume cannot be varied
    1. Solution :Check the 100k linear pot, if faulty replace it.

      Details : Refer Cable FP2.1

  • Print out is blank
    1. Solution :Flex cable of the printer not connected properly to PCB 12027 (ident J2). Check and connect it properly.

      Details : Refer Document A1

    2. Solution :If Flex cable is connected properly then check the printer (check with new printer).

      Details :Refer Document A1

    3. Solution Change PCB 12027 - Printer PCB.

      Details :Refer reference document for making changes in the unit and and Refer Document A1

  • Printer paper rolls in the opposite direction
    1. Solution :Flex cable of the printer not connected properly to PCB 12027 (ident J2). Check and connect it properly.

      Details : Refer Document A1

    2. Solution :If Flex cable is connected properly then check the printer (check with new printer).

      Details :Refer Document A1

    3. Solution: Change PCB 12027 - Printer PCB.

      Details :Refer reference document for making changes in the unit and Refer Document A1

  • QRS volume not coming
    1. Solution :Check the 100k linear pot, if faulty replace it.

      Details :Refer Cable FP2.1

    2. Solution :Check buzzer (B20) on PCB 9013 (ident Buzz 1) Control card, if faulty replace it.

      Details :Refer Image 1

  • Shock is not delivered
    1. Solution :Check continuity of U plate connection from Apex handle to PCB 12026 (ident CON6), If faulty replace it.

      Details :Refer Cable FP8 and Image 6

    2. Solution :Check continuity of U plate connection from Sternum handle to PCB 12026 (ident CON5), If faulty replace it.

      Details :Refer Cable FP7 and Image 6

    3. Solution :Check the corrosion on adult and child plate of the handle. If they are corroded replace them. The corrosion can affect the accuracy of shock delivery.

      Details :NA

    4. Solution :Change FRC cable connecting PCB 9013 or (ident CN9) to PCB 11006 (ident J2).

      Details :Refer Cable CC1 and Image 4

  • SpO2 waveform is flat
    1. Solution :Check continuity of Spo2 connector from PCB 9013 to PCB 14015, If faulty replace it.

      Details :Refer Cable SP2.2

    2. Solution :Check continuity of Spo2 connector from PCB 14015 to side panel of the unit, If faulty replace it.

      Details :Refer Cable SP2.1

    3. Solution :Change PCB 14015 - SpO2 card.

      Details : Refer reference document for making changes in the unit

  • Unit does not turn 'ON' on Mains
    1. Solution :Check if fuse is faulty or fuse holder is broken (On the back panel). If faulty replace it. Fuse should be 8 A. With voltage rating of 250v.

      Details :NA

    2. Solution :Check adaptor output, on PCB 9013(ident CN3). It should be 17v. Check with respect to pin no 1(output) and 3(ground). If it is less than 17v then, replace the adaptor.

      Details :Refer Image 1

  • Unit is not charging
    1. Solution :Change FRC cable connecting PCB 9013 (ident CN9) to PCB 11006 (ident J2).

      Details :Refer Cable CC1 and Image 4

    2. Solution :Check if the unit is charging using 'Charge' button on the Apex handle. If yes then the Blue charge switch on the front panel may be faulty, replace it.

      Details :Refer Cable FP10 and Refer Document A2